Table 2 Pulse sequence parameters of MR-guided in-bore biopsy

From: MRI-guided in-bore biopsy of the prostate – defining the optimal number of cores needed

 

sag. T2w HASTE

ax. T2w TSE

sag. T2w TSE

ax. DWI b0, 1500

paraax./ parasag. T2w TSE

paraax. TruFISP

TR (ms)

1400

7500

7500

4640

6500

690

TE (ms)

101

101

101

78

96

1.8

Flip angle (degrees)

126

160

160

180

160

50

Fat saturation

No

No

No

No

No

Yes

Slice thickness (mm)

3

3

3

3

3

3

Gap (mm)

0

0

0

0

0

0

Matrix

320 × 256

320 × 288

320 × 288

112 × 112

320 × 310

256 × 175

FOV (mm)

300 × 300

220 × 220

240 × 240

220 × 220

300 × 300

300 × 300

Acquisition time (min: sec)

0:22

2:52

2:52

3:58

0:59

0:22

  1. Sag. = sagittal, HASTE = Half-Fourier Acquisition Single-shot Turbo spin Echo, ax. = axial, TSE = Turbo spin echo, DWI = Diffusion weighted imaging, paraax. = paraxial, parasag. = parasagittal, TruFISP = True fast imaging with steady-state free precession, TR = Repetition Time, TE = Time to echo, FOV = Field of view